3-8 Judging “Above” and “Below” (Limit Function)
-43-
Judging by deviation values
The function setting for deviation value judgment is “
.”
Judging by deviation values using the method of placing actual samples on the
balance
Set a reference, upper limit, and lower limit values by weighing samples.
Check that the limit function is set for judging by deviation values (refer to the previous section
“Limit function setting”).
1
Set to the mode of using actual samples.
<Press and hold>
Press and hold the [Set] key. After “
” is
displayed, release the finger.
2
Weigh the reference weight sample.
Place the sample serving as the reference value
and then press the [Function] key.
3
Weigh the lower limit sample.
Place the sample serving as the lower limit and
then press the [Function] key.
For the case of one point setting, this operation
completes the setting saving. Go to step 5.
For the case of two point setting, “
” is
displayed.
4
Weigh the upper limit sample.
Place the sample serving as the upper limit weight
on the pan and then press the [Function] key.
The judgment value is saved and then the balance goes
back to normal display.
5
Weigh.
Place a judgment target sample on the pan.
The OK, LO, or HI result judged from the upper and
lower limit values is displayed with the “
” indication.
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......