6-3 Communication Data and Commands
-69-
- When expression units are changed or switched to display of parts counting, percentage
weighing, or unit converting, the decimal place is changed (Refer to “
Sample
communication formats” (P. 70)).
[U1, U2] (two characters)
Indicates the unit used to show numeric data.
U1 U2
Code
Meaning
Balance
indicator
(SP) G 20H
47H
gram
g
K G
4BH
47H
Kilogram
kg
C T
43H
54H
carat
CT
O Z
4FH
5AH
ounce
L B
4CH
42H
pound
O T
4FH
54H
troy
ounce
D W
44H
57H
pennyweight
G R
4BH
52H
Grain
Bottom
right
grain
T L
54H
4CH
tael
(Hong
Kong)
T
L
54H
4CH
tael (Singapore, Malaysia)
Top right
T
L
54H
4CH
tael (Taiwan)
Middle right
M O
4DH
4FH
momme
mom
t o
74H
6FH
tola
P C
50H
43H
Pieces
Pcs
(SP) % 20H
25H
Percentage
%
(SP)
#
20H
23H
Computation results, numbers, etc.
#
[S1] (one character)
Indicates the judgment result when the limit function is used.
S1 Code
Description
Remarks
L
4CH
Below (LO)
1- or 2-point scale
G 47H
Appropriate
(OK)
H 48H
Above
(HI)
T
54H
Cumulative value
Data type
U 55H
Unit
weight
(SP)
20H
No judgment result or no data type specified
d 64H
Gross
Содержание CJ-E series