5-2 Minimum Readability Setting
-53-
5-2 Minimum Readability Setting
Use this function to set the minimum readability. The larger the minimum readability becomes,
the less the balance is affected by external influences. In addition, it takes less time for the balance
reading to become stable.
Each unit has different minimum readability.
1
Set to the function setting mode.
(Refer to “Section 2-5: Function Setting Basics (P.
16).)
Press and hold the [Function] key. After “
”
is displayed, release the finger.
2
Select the minimum readability.
Press the [Function] key several times to select “
.”
Select from 1 through 5 by pressing the [Zero/Tare]
key.
3
Save the setting.
Press the [Set] key.
The balance goes back to normal display.
参 考
To set the minimum readability of unit B, select “
” in step 2.
You can also set the same unit for unit A and unit B, and set different minimum readability, so
that unit A and unit B can be used to switch the minimum readability.
Minimum readability example
Setting
value
CJ-220E~820E CJ-2200E~8200E
CJ-15KE
g kg ct
mom
g kg ct
mom
g kg ct
mom
1 0.01
0.00001 0.05
0.005 0.1 0.0001
0.5
0.05
1 0.001 5 0.5
2 0.02
0.00002
0.1 0.01 0.2 0.0002
1
0.1 2 0.002 10 1
3 0.05
0.00005
0.2 0.02 0.5 0.0005
2
0.2 5 0.005 10 2
4 0.1
0.0001
0.5
0.05 1 0.001 5
0.5 10 0.01 10 5
5 0.2
0.0002
1 0.1 2 0.002 10
1 10 0.02 10 10
Reference
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......