Adjusting the Scale
-48-
4-1 How to Calibrate
To calibrate a balance is called span adjustment. Be sure to perform span adjustment for highly
accurate measurements.
An electronic balance is influenced by the acceleration of gravity. For this reason, you should
calibrate your balance every time you relocate it. You should also calibrate it after a long time of
use or when it does not indicate correct values.
注 意
To adjust the span, wait at least 15 minutes after the balance is powered on.
参 考
Use a weight for calibration that weighs 50% of the weighing capacity or heavier. To calibrate
more accurately, use a weight that is equivalent to the weighing capacity.
Please contact us if you wish inquire about or place an order for calibration weights.
1
Set to the function setting mode.
(Refer to “Section 2-5: Function Setting Basics (P.
16).)
Press and hold the [Function] key. After “
”
is displayed, release the finger.
2
Set to the span adjustment mode.
Press the [Function] key several times to select “
.”
Press the [Zero/Tare] key several times to select
“
.”
3
Save the setting.
Press the [Set] key.
The balance goes back to normal display.
4
Start span adjustment.
<Press and hold>
Keep holding down the [Function] key until “
” is displayed.
5
Start zero-point adjustment.
Releasing the finger changes the display to the
blinking of “
,” starting zero-point
adjustment.
Caution
Reference
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......