3-3 Measuring Percentage
-23-
3-3 Measuring Percentage
With respect to the reference sample weight, the weight of a sample is shown in percentage.
A reference sample weight can be set by weighing an actual sample (setting a reference
weight by weighing an actual sample) or entering a value (
setting a reference weight by
entering a value
).
Setting a re
f
erence weight by weighing an actual sample
1
Set to the function setting mode.
(Refer to “Section 2-5: Function Setting Basics (P.
16).)
Press and hold the [Function] key. After “
”
is displayed, release the finger.
2
Select “Percentage Weighing.”
Press the [Zero/Tare] key several times to select
“
.”
3
Save the weighing mode.
%
Press the [Set] key.
The “percent weighing” mode is set, displaying “%.”
4
Start measuring the reference weight.
<Press and hold>
Press and hold the [Function] key. After “
” is displayed, release the finger.
The previously-saved reference sample weight blinks.
5
Save the reference weight.
Put the reference sample on the balance and then
press the [Function] key.
The reference weight is saved.
6
Weigh.
The display indicates the percentage of the sample
with respect to the reference sample weight.
Pressing the [Function] key toggles the display between
percentage weight and total weight.
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......