3-8 Judging “Above” and “Below” (Limit Function)
-41-
7
Save the setting.
Press the [Set] key.
Then, set the values for judgment (limit values) to perform measurement.
Judging by absolute values
The function setting for absolute value judgment is “
.”
Judging by absolute values using the method of placing actual samples on the
balance
Set upper and lower limits by weighing samples.
Check that the limit function is set for judging by absolute values (Refer to the previous section
“Limit function setting”).
1
Set to the mode of using actual samples.
<Press and hold>
Press and hold the [Set] key. After “
” is
displayed, release the finger.
2
Weigh the lower limit sample.
Place the sample serving as the lower limit and
then press the [Function] key.
For the case of one point setting, this operation
completes the setting saving. Go to step 4.
For the case of two point setting, “
” is
displayed.
3
Weigh the upper limit sample.
Place the sample serving as the upper limit weight
on the pan and then press the [Function] key.
The judgment value is saved and then the balance goes
back to normal display.
4
Weigh.
Place a judgment target sample on the pan.
The OK, LO, or HI result judged from the upper and
lower limit values is displayed with the “
” indication.
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......