6-3 Communication Data and Commands
-73-
(2) Set output control command
C1
C2
Code (C1)
Code (C2)
Description
Response
O 0 4FH
-{}-30H Stop
output.
O
1
4FH
31H
Output continuously at all times.
O 2 4FH
32H
Output continuously if stable
(Stop output if unstable).
O 3 4FH
33H
Output once when the [Print]
key is pressed (whether the
balance is stable or unstable).
O 4 4FH
34H
Output once when the balance is
stable. Output when a sample is
lifted to cause the display to
indicate a value below zero, and
then another sample is placed to
make the balance stable.
O 5 4FH
35H
Output once when the balance is
stable. Stop output when
unstable. Output once when the
balance is stabilized again (the
output includes zero) even if it
is not reloaded.
O 6 4FH
36H
Output once when the balance is
stable. Output continuously
when unstable. Output is
stopped after a single output
when the balance is stable even
if it is not reloaded.
O 7 4FH
37H
Output once when the [Print]
key is pressed if the balance is
stable.
O 8 4FH
38H
Output
once
immediately.
O
9
4FH
39H
Output once after stabilized.
A00: Successful
completion
E01: Command
error
- Commands O0 to O7 have the same workings as the output control set by the function setting.
- Commands O8 and O9 are used to request data from the balance.
- Once executed, O0 to O7 commands are held. However, the status is reset to the function
setting when the balance is turned on again.
- When either an O8 or O9 command is executed, it returns to the state of “O0”.
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......