Troubleshooting
-78-
7-1 Error Messages
Message Cause
Remedy
- The weight of the sample is over the
weighing capacity
- The number of digits in the addition result
or calculation result went over the number
that can be displayed.
- Unload the sample to weigh it in
some portioned-out measurements.
- Replace the tare with a lighter one.
-
If the error message does not
disappear even when nothing is
placed on the pan, mechanical parts
may have failed. Contact our Sales
Office or Technical Service Division.
- First, clear the addition result. Then
execute addition again.
-
The coefficient used in unit
converting is too small. Set a greater
coefficient.
The minus-value load exceeded the lower
limit.
- The pan or the pan base may not be
properly set. Check them, paying
attention to whether they are in
contact with an external object.
-
If the error message does not
disappear even if the pan and the pan
base are properly set, mechanical
parts may have failed. Contact our
Sales Office or Technical Service
Division.
The reference weight used during span
adjustment with an external weight is far
less than 50% of the weighing capacity.
For span adjustment with an external
weight, use a weight that weighs as
close to the weighing capacity as
possible.
An error over 1.0% was detected in span
adjustment with an external weight, or the
balance failed.
For span adjustment with an external
weight, check that a correct weight is
placed and that no objects other than
the weight are placed. Then, execute
span adjustment again.
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......