3-8 Judging “Above” and “Below” (Limit Function)
-39-
A limit value can be entered in the following two ways:
- Putting actual samples on the balance: By weighing a sample on the balance, save the weight.
- Entering values: Set a value using the keys.
Detailed function setting
In the function setting function, the limit function can be set in detail.
When the function setting “
” is “
” or “
,” pressing the [Function] key can set the
following: Set these items as required.
Condition
1:
Always
judge.
2: Judge only when the balance is stable.
Range to Cover
0: Detect when the limit is exceeded by more than five
divisions.
1: Detect both when the limit is exceeded and when it is not
reached.
Point Scale
1: 1-point scale (OK and LO are judged.)
2: Upper and lower limits are set (HI, OK and LO are
judged).
Judge by
1: Judge by absolute values.
2: Judge by deviation values.
Buzzer for LO
0: The buzzer is not beeped for rank LO.
1: The buzzer is beeped for rank LO.
Buzzer for OK
0: The buzzer is not beeped for rank OK.
1: The buzzer is beeped for rank OK.
Buzzer for HI
0: The buzzer is not beeped for rank HI.
1: The buzzer is beeped for rank HI.
注 意
- Separate limit values can be saved for each weighing mode. However, limit values
of both absolute and deviation values cannot be saved in the same weighing mode.
- Limit values can be set only in measurement mode display (Cannot be set in other
mode display such as when a cumulative value is displayed).
- Before setting limit values, as required, adjust the zero-point or set a tare range.
- If the limit value entries are not lined up in the order of magnitude, three “
” will be
lit. Enter the values again.
Caution
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......