Various Measuring Methods
-44-
Judging by deviation values using the method of entering values
Set a reference, upper limit, and lower limit values by entering values using the keys.
Check that the limit function is set for judging by deviation values (Refer to the previous section
“Limit function setting”).
For making judgments by deviation values using the method of entering values, the lower and
upper limit values to be entered are their differences with respect to the reference weight.
For example, when a judgment is made with an upper limit of 1,050 g and a lower limit of 900 g,
enter a reference weight of 1000 g, an upper limit of 50 g, and a lower limit of -100 g.
1
Set to the mode of entering values.
<Press and hold>
Press and hold the [Set] key. After “
” is
displayed, release the finger.
2
Enter the reference value.
Enter the reference value with the following steps:
1. Press the [Zero/Tare] key.
Zero blinks in the right end.
2. Select a number by pressing the [Zero/Tare]
key.
Pressing the key toggles the number between
0 and 9, and decimal point.
3. Pressing the [Function] key shifts the blinking
digit to the left, and the next lower-order digit
blinks.
4. Press the [Set] key to save the value.
3
Enter the lower limit value.
Enter the lower limit value by using the same
operation as with step 2.
For the case of one point setting, this operation
completes the setting saving. Go to step 5.
For the case of two point setting, “
” is
displayed.
4
Enter the upper limit value.
Enter the upper limit value by using the same
operation as with step 2.
The judgment value is saved and then the balance goes
back to normal display.
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......