Various Measuring Methods
-30-
turned off.
- The acceptable range of water temperature is
0 to 99.9°C.
- Pressing the [Print] key can cancel the setting.
6
Save the water temperature.
Press the [Set] key.
7
Hang only the cage to zero the weight
readout.
Hang only the cage on the underfloor weigher
fixture.
Press the [Zero/Tare] key.
The readout becomes zero.
8
Measure the weight of the sample.
Place the sample on the cage.
After the weight display is stable, press the [Set]
key to save the weight of the sample in the air.
When the weight is saved, “
” is displayed in the
lower left of the display.
Measuring a sample by placing it on the pan is also
allowed.
9
Set the liquid container.
Set the container with water under the balance and
submerge only the cage (do not place the sample).
Press the [Zero/Tare] key to zero the readout.
This is to remove any residual error due to the cage.
10
Place the sample on the cage.
Put the sample on the cage and submerge the
entire volume.
After the weight display is stable, press the [Set]
key.
The measured specific gravity value is displayed.
注 意
Do not allow the cage to touch the liquid container.
参 考
Press the [Set] key to return to showing the
sample’s normal weight.
Caution
Reference
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......