Site Preparation / Installation
03-914642-00:6
2-3
The Archon dimensions are: 21.5W x 17H x 21D inches; 54.6W x 43.2H x 53.3D
centimeters (see Figure 2-1), and weighs approximately 80 lbs./36.5 kilograms.
Figure 2-1
2.5
Gas Connections
Ultra high purity helium (99.999% GC/MS grade) is required for purging soil
samples and operation of the vial gripper arm. The input supply pressure must be
60-90 PSI.
The helium gas line must be connected to the bulkhead fitting labeled “Helium
Input.” Gas line must be 1/8” diameter copper tubing and fittings must be 1/8"
Swagelok®. Use a single, two-stage, high quality pressure regulator with a
stainless steel diaphragm. Consult your supplier for the proper type and size of
cylinder valves. An on-off, shut off valve, should be installed in the line to facilitate
installation. Gas line tubing should be generous. Allow enough length, in a coil, to
be able to move the Archon around to access the rear panel. Tighten and leak
test all connections using propanol and water.
Avoid any pipe thread connections in the supply line. If any fitting requires a
sealant, use an instrument-grade Teflon tape.