System Setup
03-914642-00:6
6-5
Note: If calibration test fails go to Section 6.4.2.
Note: Before beginning refer to the Default Operation Value Table on page 6-15.
6.4.2 Auto Calibrate
The "Auto Calibrate" program allows the Archon to automatically establish the
correct coordinate position settings for removal and replacement of sample vials
in the sample tray, water vial sampling station, soil/solid sampling station, and the
equilibrium ID station. These settings are essential for accurate vial / gripper
movement. After the Archon has been installed to the purge and trap, and
before
running the first sample
, run the "Calibration Test" program. The program
should also be run after ANY service to the vial gripper assembly or if ANY
gripper movement error message has been displayed. Also, if the Archon has
been relocated for some reason, run the "Calibration Test " program.
Note: Once Auto Calibrate is initiated, all other calibration steps must be
performed to assure the unit is properly aligned.
Locate the Calibration Bar (P/N DY-505894-00) and Calibration Pin (P/N DY-
505893-00) shipped with your instrument.
Note: To order a replacement Calibration Bar and Calibration Pin, order Kit
Number DY-505419-90.
To run the "Auto Calibrate" program, press the "System” key, from the key
pad. The *System Setup* menu will be displayed. Using the “2” or “8” key scroll
to the "Calibration" menu line and press ENTER. Scroll up one position using
the “8” key to the “Auto Calibrate" menu line. Press ENTER. The following
menu is displayed:
Insert Bar in Tray
10 and 17. Press
Enter when ready or
- to Abort.
Insert the Calibration Bar into the sample tray in positions 10 and 17.
Press ENTER.
The following message will appear:
Insert Bar Sensor in Gripper.
Press any key to
Continue.
Slide the bar sensor device up into the vial gripper. Note: The end with the silver
rod protruding through the device goes DOWN. Press any key to begin the
calibrate program. The vial gripper will now move to the Calibration Bar in the
sample tray, then to the water vial sampling station, equilibrium ID station and
then the soil / solid sampling station.
After completion of the program, the following message will appear:
Remove Bar Sensor in Gripper.
Press Any key to Continue.