Lines attached to the upper soil valve must be installed with a Valco
stainless nut, P/N 28-694501-00 and a set of Swagelok stainless
ferrules, P/N 28-693996-00 and 28-693997-00.
Perform a soil purge leak check by checking and balancing flows at
the soil probe gas outlet bulkhead Swagelok fitting. Record flow rates.
Also, check flow at the vent fitting on the host purge and trap. (Flow
should be within
2 ml per minute.) Record flow rates.
Check vial calibration settings. The vial MUST move freely in and out
of the tray without resistance. Refer to Section 6.4.
Permanant marker pens are recommended to label sample vials.
Otherwise, no more than two labels should be applied to the vials.
The waste bottle or drain
MUST
be placed at a location which is equal
to or lower than the base of the Archon.
Make sure the sample needle in the sprage tube does not sit centered
or flush on the glass frit. The needle should touch the side of the
glass and sit slightly above the frit.
Review Sections 3, 4, 5, and 6.
Run a water method program (Section 5.4.1).
Run a soil method program (5.4.2).
Note:
Use 7-10 mLs of clean H
2
0 for (Install Samples) on soil.