Intended Use
D292144.2.fm
vi
Operating Environment
The TM-500 is designed to be operated in a temperature and humidity controlled, light, industrial
setting. It requires an ISO Class 9 clean room environment.
The machine should be installed on a flat, dry, stable surface in a well lit area (ambient lighting of
200 to 300 Lux (Lumens/m2).
The recommended climate is between 5 - 75% non-condensing humidity with a room tempera-
ture between 0 - 50 degrees Celsius.
Note
: Although all of the components used on the machine will withstand the temperature
range of 0 to 50 degrees Celsius, such temperatures may decrease the life of some of the
components.
The intended electrical environment is Pollution Degree 2 and Over Voltage Category II.
Misuse
The TM-500 weighs approximately 1,200 pounds (545 kg) and requires a forklift or other lifting
equipment to be raised off the floor. It’s built so its legs can be raised to engage four wheels on its
bottom. When the wheels are engaged it can be rolled by a minimum of two people to a new
location. Operators are cautioned to follow the instructions and illustrations for safely transporting
the TM-500. Failure to do so could result in injury or damage to the machine.
Installation instructions clearly state that the TM-500 should only be installed or operated on a
level floor to prevent toppling. When positioned for operation, the four legs should be lowered
and adjusted so the machine is level and immobile to ensure safe and stable operation.
The TM-500 has been designed to make operation impossible during service, maintenance or at
any point when the machine’s cabinet doors are open. Do not block, disable or attempt to over-
ride the cabinet door sensors.
The user is protected from moving parts and exposure to objects being ejected under pressure
by 16 standard wire gauge sheet metal and 3/8 inch clear, polycarbonate enclosures. Door sen-
sors ensure all operation automatically stops If the top enclosure doors are opened. No attempt
should be made to remove these enclosures or override the door sensors.
The user is protected from the TM-500’s heat sealer by a metal guard. Operators are cautioned
not to touch the heat seal guard or to try to reach underneath the guard while the heat sealer is in
operation.
The TM-500 air supply should be adjusted to operate at a maximum working pressure not
exceeding 7.6 bar (110 psi). The recommended source pressure is 100 psi. The maximum per-
missible source pressure should not exceed 9 bar (130 psi). Failure to set the within these limits
could result in a failure in the pneumatic system which could lead to injury.
The Ionizing Fan air regulator should be adjusted to operate at a maximum working pressure not
exceeding 2.8 bar (40 psi). Failure to set it within these limits could damage the fan.
Содержание TM-500
Страница 2: ...European Representative QNET BV Hommerterweg 286 6436 AM Amstenrade The Netherlands...
Страница 17: ...Contact Information D292144 2 fm xiv...
Страница 21: ...Table of Contents D292144 3a fm...
Страница 47: ...2 14 TM 500 User s Guide Dual Tray Feeder Input Output Module D292144 5 fm...
Страница 83: ...3 36 TM 500 User s Guide The Setup Screens D292144 6 fm...
Страница 187: ...7 14 TM 500 User s Guide Ethernet Networks D292144 10 fm...
Страница 254: ...751 Summit Avenue Mankato MN 56001 507 387 2039 FAX 507 387 2257 www vtekusa com Email info vtekusa com...