IMPORTANT
Use crimping terminals for the connection and tighten screws.
Check also screws fixing the connector tighten.
Motor rotation direction is clockwise viewed from the motor side.
Refer to the arrow mark castled on the Motor flange
Be sure also to put a safety circuit such as MCCB(Molded Case Circuit
Breaker), MC(Magnetic Contactor) and THR(Thermal Relay) for the
electrical connection.
Ensure to have a correct grounding.
You have a risk of getting electrical shock in case of failure or electric
leakage.
You are recommended further to install a dedicated earth leakage breaker.
Minimize the length of the ground wire
Wire size, please determined by considering the voltage drop of the wire.
Typically, the voltage drop, please to be within 2% of the rated voltage of
the motor.
Voltage drop calculation
:
√
3
×
wire resistance (
Ω
/ km)
×
Wiring length (m)
×
motor rated
current (A)
×
10
-3
※
motor rated current, refer to "Table 3".
WARNING
Install and operate this product in compliance with the laws and
regulations relating to the safety, e.g. Fire Defense Law, Electric wiring
regulation and so on. In the country and region you use this product.
WARNING
IE3 motor is used for this pump.
Striking current tends to be high because efficiency of IE3 motor is higher
than conventional motor. Because of this consequence, there could be
momentary operation by striking current of the motor in the case of
current set rating of MCCB (Molded Case Circuit Breaker), ELCB (Earth
Leakage Circuit Breaker) and THR (Thermal Relay).
It is required to readjust setting of MCCB, ELCB and THR.
Содержание VS1501
Страница 2: ......
Страница 13: ...SK00 4390 DI 002 11 xi Fig 1 Warning Label display position...
Страница 27: ...SK00 4390 DI 002 11 7 Fig 2 VS1501 Dimensional drawing...
Страница 28: ...SK00 4390 DI 002 11 8 Fig 3 VS2401 Dimensional drawing...
Страница 36: ...SK00 4390 DI 002 11 16 Fig 5 Suction Exhaust piping connection diagram...