SK00-4390-DI-002-11
22
CAUTION
The Pump oil might deteriorate in a shorter time depending on the use.
It is recommended to replace the first Pump oil within ten days after
operation start and see how it got dirty to determine the oil replacement
cycle.
If the Pump breathes in a lot of water or the like, you should replace the
oil more frequently. If kept operation without getting rid of breathed
water, it would deteriorate lubrication of the oil and further help
corrosion of the Pump inside and result in causing a failure.
Do not store the product keeping sucked the water.
If the Pump breathed in chemical material such as acid, immediately
replace the oil as it would cause the rust during the stop in one night to
make the system e not applicable to operate.
We shall be not liable to the durability to chemical material.
You should also replace the oil if breathed in the solution to deteriorate
lubrication of the oil as it would also cause biting inside. You shall have
a risk if breathed in the solution in operation even you replaced the oil.
Caution shall be required for the operation under high pressure range.
Continuous operation one hour or more under high pressure 1000Pa or
more would increase the oil volume that are discharged as the oil mist
and cause rapid parts wear or cause a trouble such as burning.
You are recommended to control the oil level on regularly supplying the
pump oil. The maintenance cycle might become shorter.
Содержание VS1501
Страница 2: ......
Страница 13: ...SK00 4390 DI 002 11 xi Fig 1 Warning Label display position...
Страница 27: ...SK00 4390 DI 002 11 7 Fig 2 VS1501 Dimensional drawing...
Страница 28: ...SK00 4390 DI 002 11 8 Fig 3 VS2401 Dimensional drawing...
Страница 36: ...SK00 4390 DI 002 11 16 Fig 5 Suction Exhaust piping connection diagram...