Service Manual
Chapter 5
5‑17
5.6.5 RGT (Reagent Tray)
1. Explanation of Function
2. INIT
To move to the home position. (And to stop when detecting the
home sensor.)
3. OFFSET
To move for the corrective value set in the offset screen.
4. ADJUST
To input the angle to move by value, and select the direction of
rotation.
5. CW-1Pitch
To move clockwise by one bottle.
6. CCW-1Pitch
To move counter-clockwise by one bottle.
7. CCW-1.6(BOTL12) To move counter-clockwise to R2 aspiration position.
8. CW-1.6(BOTL1)
To move clockwise from R2 aspiration position to the basic
position.
9. POSITION
10. Low Speed
* Starting pulse rate
11. High Speed
* The highest speed pulse rate
12. Pulse Count
* Adjustable speed pulse count
13. Init Speed
* Pulse rate of initialization
* shows that setting is not needed.
Manual Function (RGT)