1‑6
1.3.
Panels
Fig. 1-8 System slant view
There are nine panels. ( Please refer to Fig. 1-6.)
1. Front panel
It is fixed by one screw and four catches. When removing, loosen the screw and
pull the panel towards you.
2.
Top
panel
This panel is fixed by five screws. When removing, remove front panel first, and
push over upper chassis after removing upper screws of right and left side
panels.
3. Right side panel
This panel is fixed by the upper and lower two screws and three catches.
When removing, open top cover first and remove two screws to leave ISE door
ajar. And loosen right screw of the reverse side of front panel.
4. Left side panel
This panel is fixed by the upper and lower screws (two) and four catches.
When removing, open top cover first to remove two screws. And loosen the right
side screw of the reverse side of front panel.
5. Rear panel (upper)
The upper panel in the back view. Fixed by nine screws. This panel has a fan.
6. Rear panel (lower)
The lower panel in the back view. Fixed by ten screws.
Right Side Panel
Front Panel
Top Panel
ISE Door
Left Side Panel
Right Side
Fixing Screw
Right Side Panel
Fixing Screw
Front Panel
Fixing Screw
Top Panel
Fixing Screw
Top Panel
Fixing Screw
Upper Chassis
Fig. 1-8 System slant view
Содержание Labmax 240
Страница 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Страница 12: ......
Страница 18: ......
Страница 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Страница 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Страница 43: ...Service Manual Chapter 2 2 11 2 5 Sample Dispensation Process Fig 2 6 Sample Dispensation Process 1 2...
Страница 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Страница 46: ......
Страница 110: ......
Страница 140: ......