Service Manual
Chapter 2
2‑7
As the last step of the measurement, for the 1
st
test of No. 1 sample, optical
absorption data are converted to concentration/activity data using calibration
curve. But test results will not be printed out till whole test results of No.1
sample are obtained.
The same process is repeated for the 2
nd
, 3
rd
… test items, and the 2
nd
, 3
rd
… sample.