Service Manual
Chapter 4
4‑11
3. When breaking down of this board
When this PCB is broken down, one or more of the following phenomena may occur.
1. Sample tray (STF) motion failure occurs.
Note In case that STF motor, sensor, encoder, or mother driver has trouble,
STF motion failure may occur. Checking is necessary.
2.
Sample tray number reading failure occurs.
Note
In case that ID sensor of STF has trouble, sample tray number reading
failure may occur. Checking is necessary.
3.
Sample probe (STM) motion failure occurs.
Note
In case that STM motor, sensor, encoder or motor driver has trouble,
STM motion failure may occur.
4.
Mixing pump (MP) won’t rotate.
Note
In case that MP motor or motor driver has trouble, MP motion failure
may occur.
5.
Motion failure of SP 1 (Sample pump for washing) occurs.
Note
In case that SP 1 motor sensor or motor driver has trouble, SP 1 motion
failure may occur.
6.
Motion failure of SP 2 (Sample pump for sample) occurs.
Note
In case that SP 2 motor sensor or motor driver has trouble, SP 2 motion
failure may occur.
Содержание Labmax 240
Страница 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Страница 12: ......
Страница 18: ......
Страница 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Страница 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Страница 43: ...Service Manual Chapter 2 2 11 2 5 Sample Dispensation Process Fig 2 6 Sample Dispensation Process 1 2...
Страница 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Страница 46: ......
Страница 110: ......
Страница 140: ......