5‑8
5.3 Result Print Setting
1.
Follow the procedures 1 to 5 in “5.2.1 Password setting procedure”
to open “C: U24pdev_N Prg t224com.ini”.
2.
Input 0, 1 or 2 in the part of “?” of “Print Mode = ?” in the
Section [Internal Printer].
“0” means that both ID and patient name are not printed.
“1” means that sample ID and patient name are printed.
“2” means that patient ID and patient name are printed.
3.
Save it.
4.
Shut down the system, and restart it.