3‑58
2. Action (Refer to “ASSY-16” for assembly drawing.)
The shaft of stepping motor (7) slides up or down when pulse is given.
This sliding motion is transferred to plunger (3).
Plungers (3) have shade pin which shade home position sensor (6) light at home
position. This detects home position at initialization.
- Home position is at the top. (Dispensation ending position)
At aspiration, shaft of stepping motor (7) goes down, and plunger also goes down.
At dispensation, shaft of stepping motor (7) goes up, and plunger also goes up.
If plunger will not be at the home position within the fixed time, system gives an
alarm and emergency stop.