Service Manual
Chapter 5
5‑21
5.6.9 RCT (Reaction Tray)
1. Explanation of Function
2. INIT
To move to the home position.
3. OFFSET
To move for the corrective value set in the offset screen.
4. ADJUST
To input the angle to move by value, and select the direction of
rotation.
5. CW-1/2 Pitch
To move clockwise by 1/2 pitch.
6. CW-1 Pitch
To move clockwise by one pitch.
7. CCW-1 Pitch
To move counter-clockwise by one pitch.
8. Stage-A
To move counter-clockwise by 13 pitches.
9. Stage-B
To move counter-clockwise by 21 pitches.
10. Stage-C
To move counter-clockwise by 27 pitches.
11.
POSITION
12. Low Speed
* Starting pulse rate
13. High Speed
* The highest speed pulse rate
14. Pulse Count
* Adjustable speed pulse count
15. Init Speed
* Pulse rate of initialization
* shows that setting is needed.
Manual Function (RCT)