Service Manual
Chapter 3
3‑33
8.
After removing fixing nuts (two) of
MU-1, remove MU-1.
- Lay removed MU-1 nearby lightly
with mixing base and piping.
9.
After removing fixing nuts (two) of
MU-2, remove MU-2
- Lay removed MU-2 nearby lightly
with mixing base and piping.
10.
Remove fixing screws (four) of
joining
plate.
Fig. 3-46 MU-1
Fixing Nut
Fig. 3-47 MU-2
Fixing Nut
Mixing Base
Fig. 3-48 Lamp House periphery
Fixing Screws (four)
Joining Plate
Lamp House
Содержание Labmax 240
Страница 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Страница 12: ......
Страница 18: ......
Страница 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Страница 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Страница 43: ...Service Manual Chapter 2 2 11 2 5 Sample Dispensation Process Fig 2 6 Sample Dispensation Process 1 2...
Страница 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Страница 46: ......
Страница 110: ......
Страница 140: ......