Service Manual
Chapter 1
1‑9
1.4.3. Unit Layout in the Left Side View
This drawing shows the left side view when the left side panel is removed.
Fig. 1-11 Unit layout in the left side view
1. Lamp DC PCB
PCB to supply power to light source lamp or others
2. Noise filter
Filter to eliminate noise from power supply
3. RS-232C connector
Connector to personal computer (RS-232C)
4. Main power switch
When the switch on, cooling function is on.
5. AC power input
Plug for AC power
6. Fuse
Fuses from F2 to F9 are set.
F9 (T-250V 5A)
F8 (T-250V 2A)
F7 (T-250V 3.15A)
F6 (T-250V 1A)
F5 (T-250V 1A)
F4 (T-250V 3.15A)
F3 Not connected
F2 (T-250V 2A)
The rating plate for fuse is attached to the inside of
left
side
panel.
7. SSR AC PCB
PCB to control the unit operated at AC 100V.
8. SSR
On-Off switch to control the temperature for reagent
cooling
system.
9
7
6
5
4
2
1
3
10
8
11
12