Service Manual
Chapter 3
3‑59
3.14.3 Unit Dismantling and Re-assembly
1.
Unit
dismantling
1. Turn off system power source and PC
power
switch.
2. Open top cover.
3. Remove front panel.
4. Remove piping (two) connected to
solenoid valve.
When disconnecting tubes, water in
the tube flows out. Wipe it off with
tissue
paper.
5. Loosen the lower screw fixing unit,
and remove the upper screw, and
then remove unit.
6. Remove three connectors connected to
unit.
J5301, J5302, J5512
Fig. 3-85 PWP periphery
Fig. 3-84 System front view
Fixing Screw
Joint of Piping
Joint of Piping
Fixing Screw
Содержание Labmax 240
Страница 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Страница 12: ......
Страница 18: ......
Страница 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Страница 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Страница 43: ...Service Manual Chapter 2 2 11 2 5 Sample Dispensation Process Fig 2 6 Sample Dispensation Process 1 2...
Страница 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Страница 46: ......
Страница 110: ......
Страница 140: ......