Service Manual
Chapter 3
3‑47
3.11.3. Unit Dismantling and Re-assembly
1.
Unit
dismantling
1.
Loosen fixing screws (six) of the plate,
and remove them.
2.
Remove
the
plate.
3.
The left side view shows joint part of
piping.
Remove piping for washing (two)
inserted in washing nozzles.
4.
The left side view shows manifold.
Loosen blue nut under manifold, and
remove connected piping.
Fig. 3-71 Blue Nut
Piping Joint Nut
Fig. 3-70 Piping of CWS
Manifold
Piping for washing (two)
Fig. 3-69 RCT periphery
Fixing Screws (six)
Plate