5‑34
5.7.4 Gain Screen
1. Explanation of the screen
When clicking Gain button in the bottom of the manual function screen,
the following screen is shown.
This screen is used for adjusting the gain automatically when exchanging
the
lamps.
2. The procedure for adjusting the gain
1) Click Read All button and check that the values are displayed.
2) Click Offset button in the row of All CH Adjust to set Offset automatically.
3)
Clicking Parameter button, the dialog box is shown.
Auto Gain screen
Auto Gain Parameter screen
Input 1000.