3‑32
3.7.3. Unit Dismantling and Re-assembly
1. Unit dismantling
1.
Turn off system power source and PC
power switch.
2.
Open top cover and push over upper
chassis gently to give a tilt.
- Make sure to remove fixing screws
on left and right panels in advance.
- Confirm the “click” sound made by
a stay when fixed.
3.
Remove front panel first, and then
upper panel.
- Refer to Chapter 1, page 1-5.
4.
Remove the plate.
5.
Holding RTM arm, rotate and move it
to reagent tray.
6.
Remove cuvette covers (two).
7.
After taking off nut fixing nozzle plate
of reaction cuvette washing
mechanism, remove nozzle plate.
- Put removed nozzle plate nearby
lightly with piping as it is.
Fig. 3-44 Reaction Tray periphery
Fig. 3-45 Reaction Cuvette Washing Mechanism
Nozzle Plate
Fig. 3-43 System front view