2‑4
2.2.
Analysis Process and Optical Measurement Point
Fig.2-3 Analysis Process Diagram
<
Explanation
>
S
:
Sample dispense
R1
:
R1 dispense
R2
:
R2 dispense
M
:
mixing
<
Analysis process
>
Cycle after start
Function
Optical
measurement
6, 7, 8
Water blank measurement
1, 2, 3
11 R1
dispense
6
17 Sample
dispense
12
17, 18
Mixing (R1 + sample)
12, 13
37 R2
dispense
32
38, 39
Mixing (R2 + reaction liquid)
33, 34
53 Mixing
before
measurement
48
60
Reaction waste aspiration
55
* For 1 regent system only, there is not R2 dispensation.
Main‑reding ponits
Sub‑reading points
This figure shows 2 points end reaction curve.
measuring point
15 second/div.
The number shows, cycle No.
1.5min
ABS
Water blank
Cuvette washing
5min
5min
Содержание Labmax 240
Страница 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Страница 12: ......
Страница 18: ......
Страница 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Страница 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Страница 43: ...Service Manual Chapter 2 2 11 2 5 Sample Dispensation Process Fig 2 6 Sample Dispensation Process 1 2...
Страница 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Страница 46: ......
Страница 110: ......
Страница 140: ......