Service Manual
Chapter 4
4‑21
3. When breaking down of this board
When this PCB is broken down, one or more of the following phenomena
may
occur.
1. Sample tray (STF) rotation error comes out, or rotating sound becomes louder.
2.
Error comes out in up/down motion of sample probe (STM), or motion sound
becomes
louder.
3. Error comes out in up/down motion of mixing unit-1 (MU-1), or motion sound
becomes louder.
4. Error comes out in up/down motion of mixing unit-2 (MU-2), or motion sound
becomes louder.
5. Error comes out in up/down motion of reagent probe (RTM), or motion sound
becomes louder.
6.
Error comes out in up/down motion of reaction cuvette washing station (CWS), or
motion
sound becomes louder.
7. Mixing pump (MP) does not rotate, or rotating sound becomes louder.