Chapter 10
|
Maintenance
184
|
Orion AquaMate Spectrophotometer User Manual
Thermo Scientific
Replacing the Tungsten Lamp
This procedure is for the Orion AquaMate 7000 Vis spectrophotometer.
Warning: This lamp gets very hot during operation. Before removing the lamp, turn off the
instrument and allow the lamp to cool for 10 minutes.
Warning: Turn off and unplug the instrument before removing the cover of the lamp
compartment.
To replace the tungsten lamp:
1.
Turn off and unplug the instrument from the wall outlet or power strip.
2.
Make sure the lamp has cooled for at least 10 minutes.
3.
Loosen the captive screw on the lamp door by rotating it counterclockwise about ¼ turn.
4.
Raise the lamp door.
Captive screw on lamp door
Aligning the lamp assembly
Thumbscrew
Lamp assembly
Tab on cable connector