Chapter 8
|
Absorbance, % Transmittance and Concentration Measurements
Thermo Scientific
Orion AquaMate Spectrophotometer User Manual
|
153
b.
Place the sample in the appropriate position on the vial holder carousel. If needed,
use the Sample Positioner keypad to set the vial location. Press the Measure
Sample function key. Repeat for each sample.
5.
When all of the samples have been measured, press the Save Data function key and
follow the onscreen prompts to save the sample measurements to the USB memory stick.
Recalling an Existing Advanced A-%T-C Method
1.
Press the test key to access the Test Types menu.
2.
Press the ▲or ▼ key to highlight Advanced A-%T-C and press the enter key.
3.
To recall an existing standard curve, press the Stored Tests function key.
4.
Press the ▲or ▼ key to highlight Load internal test or Load test from USB drive and press
the enter key.
5.
Press the ▲or ▼ key to highlight the desired test and press the enter key.