Chapter 2
|
Spectrophotometer Basics
18
|
Orion AquaMate Spectrophotometer User Manual
Thermo Scientific
Sample Positioner Setting for the Vial Holder Carousel
The instrument allows the use of various vial holder carousels to take measurements. When
selecting the test parameters, the Sample Positioner setting determines how the instrument will
work with the installed vial holder carousel to obtain readings in the test measurement screen.
The Sample Changer keys on the instrument keypad are also be used for certain settings.
Sample Positioner Settings
Setting
Description
Manual 3
This setting is only used in the Water Analysis test type.
Use the Sample Changer keys to position the carousel with the desired vial holder in the
beam pathway for measurement. Positions B, 2 and 4 are selectable.
Auto 6
Measure one blank and five samples without changing vials on the carousel.
The instrument automatically measures the blank and advances the carousel to measure
the remaining samples in positions 1 through 5.
Auto 3
Measure one blank and two samples without changing vials on the carousel.
The instrument automatically measures the blank and advances the carousel to measure
the samples in positions 2 and 4.
Manual 6
Measure the blank and up to five samples on the carousel.
Use the Sample Changer keys to position the carousel to the appropriate position for
sample measurement. Place the blank vial in the B position and samples in position 1-5.
Regardless of the carousel position, the carousel automatically goes to the B position and
measures the blank when the Measure Blank function key is pressed. Use the Sample
Changer keys to select different positions for samples.
1-Cell
Platform
Measure the blank and all samples using one position on the carousel.
Place the blank in the holder, measure the blank, remove the blank, place a sample in
the holder, measure the sample, remove the sample and repeat for all samples. The
Sample Changer keys do not function is this setting.
Note: When the carousel is installed and Manual 6 is selected, the instrument always considers
the vial in the B position as a blank. This means that even after measuring the blank the first
time, samples can only be placed in positions 1 through 5.