Chapter 5
|
Orion AQUAfast Reagent Chemistry Instructions for Orion AquaMate
Thermo Scientific
Orion AquaMate Spectrophotometer User Manual
|
87
AC4P55 Manganese High Range Powder Pack Test
Periodate Oxidation Method
0.1 – 18 mg/l Mn (High Range)
1.
Load and run the AC4P55 method.
2.
Fill a clean AQUAfast 24mm round vial, Cat. No. AC2V24, with 10 ml of sample. Close the
vial tightly with the cap. Wipe the exterior of the vial.
3.
Place the vial into the holder in the sample chamber and close the chamber door.
4.
Press the Measure Blank function key to measure the blank.
5.
Open the sample chamber door and remove the vial from the holder.
6.
Add the contents of one Manganese Citrate Buffer Powder Pack straight from the foil into
the vial.
7.
Close the vial tightly with the cap and swirl or invert several times to mix the contents.
8.
Add the contents of one Sodium Periodate Powder Pack straight from the foil into the vial.
9.
Close the vial tightly with the cap and swirl or invert several times to mix the contents. Wipe
the exterior of the vial.
10.
Wait for a reaction period of 2 minutes.
11.
Place the vial into the holder in the sample chamber and close the chamber door.
12.
Press the Measure Sample function key to display the result in mg/l manganese.
Notes:
•
This test is applicable for the determination of soluble manganese in water and
wastewater.
•
Highly buffered water samples or extreme pH values may exceed the buffering capacity of
the reagents and requires sample pre-treatment. If samples were acidified for storing,
adjust the pH between 4 and 5 with 5 mol/l (5 N) sodium hydroxide before test. Do not
exceed pH 5, as manganese may precipitate.
•
Interferences:
Interfering Substance
Interference Level
Calcium
Greater than 700 mg/l
Chloride
Greater than 70,000 mg/l
Iron
Greater than 5 mg/l
Magnesium
Greater than 100,000 mg/l
Содержание Orion AquaMate UV-Vis
Страница 2: ......
Страница 7: ......
Страница 128: ...Chapter 6 Standard Curve Test Menu 128 Orion AquaMate Spectrophotometer User Manual Thermo Scientific ...
Страница 135: ...Chapter 7 Wavelength Scanning Test Menu Thermo Scientific Orion AquaMate Spectrophotometer User Manual 135 ...
Страница 212: ......