Chapter 5
|
Orion AQUAfast Reagent Chemistry Instructions for Orion AquaMate
Thermo Scientific
Orion AquaMate Spectrophotometer User Manual
|
111
AC4P82 Sulfate Powder Pack Test
Barium Sulfate/Turbidity Method
5 – 100 mg/l SO
4
1.
Load and run the AC4P82 method.
2.
Fill a clean AQUAfast 24mm round vial, Cat. No. AC2V24, with 10 ml of sample. Close the
vial tightly with the cap. Wipe the exterior of the vial.
3.
Place the vial into the holder in the sample chamber and close the chamber door.
4.
Press the Measure Blank function key to measure the blank.
5.
Open the sample chamber door and remove the vial from the holder.
6.
Add the contents of one Sulpha 4/F10 Powder Pack straight from the foil into the vial.
7.
Close the vial tightly with the cap and swirl or invert several times to mix the contents. Wipe
the exterior of the vial.
8.
Wait for a reaction period of 5 minutes.
9.
Place the vial into the holder in the sample chamber and close the chamber door.
10.
Press the Measure Sample function key to display the result in mg/l sulfate.
Notes:
•
If sulfate ions are present a cloudy solution will appear.