Chapter 9
|
Performance Verification
Thermo Scientific
Orion AquaMate Spectrophotometer User Manual
|
177
5.
Press the Start Test function key.
6.
The results will indicate pass or fail for each wavelength.
If the test fails, follow these guidelines:
•
Repeat the test twice to verify that the test is failing consistently.
•
Make sure all the filters being used are specifically designed to measure stray light at the
specific wavelengths.
•
Verify that the filters are placed in the correct positions.
•
Contact technical support for more troubleshooting advice.