Chapter 7
|
Wavelength Scanning Test Menu
134
|
Orion AquaMate Spectrophotometer User Manual
Thermo Scientific
Collecting a Baseline Scan and Scanning a Sample
Note: If a 3-position or 6-position vial holder carousel is in use, the blank must be placed in the
B position. The instrument always uses the B position to collect the baseline.
Once the parameters have been set for the scan:
1.
Press the Run Test function key.
2.
Prepare the blank and sample that will be scanned.
3.
Place the blank in the B position on the vial holder carousel and press the Collect
Baseline function key.
4.
While the instrument is taking the baseline, a “Collecting baseline” message will be shown
indicating the progress of the baseline scan. After the baseline is measured, the message
will update to “Baseline Collected”. Remove the blank from the sample compartment.
5.
Place the sample in the appropriate position on the vial holder carousel. If needed, use the
Sample Positioner keypad to set the vial location.
6.
Press the Measure Sample function key.
7.
When the scan is complete, a graph of the scanned measurement will be displayed. Press
the Tabular/Graph function key to switch between tabular and graphical displays.