Chapter 8
|
Absorbance, % Transmittance and Concentration Measurements
150
|
Orion AquaMate Spectrophotometer User Manual
Thermo Scientific
Using the Advanced A-%T-C Test Method
Use the Advanced A-%T-C mode to obtain advanced concentration measurements. Methods,
data and factors can be saved in this application. This advanced concentration mode is useful
for very simple comparisons that do not require a standard curve. For better accuracy, use the
The Advanced A-%T-C mode measures concentration using either a standard or factor to
convert absorbance readings to concentration units. When using a standard, specify the
concentration of the standard and measure its absorbance. When using a factor, specify the
factor and concentration units. The steps for taking measurements in the two modes are similar,
the only difference is if a standard is measured or a factor is entered.
Setting the Parameters for Advanced A-%T-C Method
1.
Press the test key to access the Test Types menu.
2.
Press the ▲or ▼ key to highlight Advanced A-%T-C and press the enter key.
3.
Press the ▲or ▼ key to highlight Measurement Mode and press the enter key. Press the
▲or ▼ key to highlight Conc/Std or Conc/Factor and press the enter key.
4.
Highlight and change the displayed test parameters, including Test Name, Wavelength,
Concentration of Standard, Factor, Units and Sample Positioner.