Chapter 8
|
Absorbance, % Transmittance and Concentration Measurements
Thermo Scientific
Orion AquaMate Spectrophotometer User Manual
|
151
a.
Some parameters only appear when one of the concentration modes is selected,
while others appear regardless of the selected measurement mode. See the
section for more information on the test parameter settings.
b.
Press the ▲or ▼ key to highlight More parameters… and press the enter key to
view additional parameters displayed on the second page.
5.
Press the Save Test function key and follow the onscreen prompts to save the test.
Measuring a Standard with Advanced A-%T-C Method
Once the parameters have been set for the concentration test and the measurement mode is
set to Conc/Std:
1.
Press the Run Standard function key.
2.
Prepare the blank and standard that will be measured.
3.
Use the numeric keypad to enter the concentration of the standard and press the enter key
to save the value.
4.
If Auto 6 or Auto 3 is selected for the Sample Positioner setting:
a.
Place the blank and standard in the appropriate positions on the carousel.
b.
Press the Measure Standard function key. Confirm that the blank and standard are
in the appropriate positions on the carousel and press the enter key.