Chapter 9
|
Performance Verification
Thermo Scientific
Orion AquaMate Spectrophotometer User Manual
|
169
Accessing the Performance Verification Tests
1.
Press the test key to access the Test Types menu.
2.
Press the ▲or ▼ key to highlight Performance Verification and press the enter key.
3.
Press the ▲or ▼ key to highlight the desired test and press the enter key.
Troubleshooting Checklist
If a performance verification test fails, follow the suggestions below to diagnose common
problems. If a test continues to fail after you have tried all these recommendations, follow the
troubleshooting list for the test being run (included with the description of each test).
•
Make sure to follow the instructions for the test properly.
•
Make sure filters and standards are clean.
•
Check that the sample compartment door is closed during the test.
•
Check that the sample compartment is clear of obstructions.
•
Make sure the vial holder carousel is installed properly. If the 6-position vial holder carousel
is installed, run the test once with the sample compartment door open to verify that the
carousel is moving smoothly.
•
Check that no problems are indicated by the power-on diagnostics after you turn the
instrument power OFF and then ON.
•
Make sure the lamp is ON.
•
Make sure the lamp compartment is clear of obstructions.
Warning: Do not open the lamp compartment unless the instrument power is OFF.
Warning: Do not turn the instrument power ON unless the lamp compartment is closed.