Chapter 6
|
Standard Curve Test Menu
Thermo Scientific
Orion AquaMate Spectrophotometer User Manual
|
123
Setting the Parameters for a Standard Curve
1.
Press the test key to access the Test Types menu.
2.
Press the ▲or ▼ key to highlight Standard Curve and press the enter key.
3.
Highlight and change the displayed test parameters, including Test Name, Wavelength,
Curve Fit, Number of Standards, Units and Sample Positioner.
a.
See the
section for more information on the test parameter settings.
b.
See the
Sample Positioner Setting for the Vial Holder Carousel
information on selecting the appropriate Sample Positioner setting.