Chapter 2
|
Spectrophotometer Basics
Thermo Scientific
Orion AquaMate Spectrophotometer User Manual
|
19
Three-Position Vial Holder Carousel
Corresponding Sample Changer Keys
Initialization of the Vial Holder Carousel
When the Run Test function key is pressed to start a measurement, the instrument will display
the message “Calibrating and Checking Turret, Please Wait” while it attempts to initialize the
vial holder carousel (Sample Positioner) at the blank (B) position. With the AquaMate three-
position carousel, the instrument will rotate the carousel to the default position and then rotate
the carousel back to the last selected measurement position (B, 2 or 4). This ensures that the
vial holder is properly positioned in the beam.
If a one-position vial holder carousel has been installed, the message “Error, Single Cell Holder
found. Use Single Cell Holder?” will appear. Press the Accept Change function key to continue
or install the AquaMate three-position carousel and press the Cancel Change function key.
If the six-position vial holder carousel has been installed, the instrument will initialize to the
blank (B) position. After it has initialized the carousel, the instrument will proceed to the test
measurement screen.
Note: To prevent a Fatal Error message when removing the vial holder carousel, always return
to the main menu or the test parameter menu before removing the carousel. If the carousel is
removed while in the test measurement screen, the instrument will display ““Fatal Error: 8 Press
Esc to return to main menu.” If this message is displayed, press the esc key and then press the
Run Test function key to re-initialize the vial holder carousel.