Symptom
Possible Cause
Solution
The Reflex aperture does not
match the virtual aperture.
The aperture size, shape or
orientation has been changed
manually.
Initialize the aperture. See “Initializing the
Reflex aperture” in the “Preparing your
Microscope” chapter if you are unfamiliar
with the procedure.
The computer cannot
communicate with the aperture.
Verify that the cable between the microscope
Communications connector and the
computer serial port is connected and firmly
seated.
The computer or microscope
are malfunctioning.
Contact technical support.
The message “Failed to open
communications. All
microscope and X-Y
automation has been disabled”
appears when you start
OMNIC Atlµs.
The microscope and/or optional
motorized stage controller is
turned off.
Power off all of the system components (the
microscope, stage controller, spectrometer
and computer).
Power the system on in the following
sequence:
Microscope
Stage controller, if installed
Illuminators
Quantitative Contact Alert, if installed
Computer
Spectrometer
The message “Failed to open
communications. All
microscope and X-Y
automation has been disabled”
appears when you start
OMNIC Atlµs,
continued…
The stage controller cannot
communicate with the
computer.
Verify that the cable that connects the stage
controller to the computer is connected and
firmly seated.
Verify that the stage controller is powered
ON.
Verify that the stage controller power cord is
connected to an AC source (working wall
outlet or power strip).
Verify that the cable that connects the stage
controller to the fast mapping stage is
connected.
Replace any damaged cables or cords.
continued…
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...