Thermo Scientific
124
A variety of detectors are available to meet sampling requirements
that demand flexibility in the step and sampling area size. Used in
conjunction with the Reflex aperture, they permit accurate data
collection from samples smaller than 10 µm square. Any one of the
detectors from the tables that follow can be factory-installed. Unlike
detectors in the spectrometer, microscope detectors are not
interchangeable. You can, as required by your sample, vary the
beamsplitter and light source used with the detector to collect data
over a more broad spectral range. The table that follows lists the
single element detectors that are available and provides element size,
spectral range and sensitivity specifications for each.
Detector
Element size Spectral range
Sensitivity (D*)
MCT-A*
50 µm
11,700 - 700 cm
-1
> 60 x 10
9
W
)
Hz
(
cm
2
/
1
MCT-A*
250 µm
11,700 - 750 cm
-1
> 60 x 10
9
W
)
Hz
(
cm
2
/
1
MCT-A
250 µm
11,700 - 600 cm
-1
> 42 x 10
9
W
)
Hz
(
cm
2
/
1
MCT-B
250 µm
11,700 - 450 cm
-1
> 7 x 10
9
W
)
Hz
(
cm
2
/
1
InGaAs
250 µm
12,000 - 3800 cm
-1
> 80 x 10
9
W
)
Hz
(
cm
2
/
1
Once you have selected the microscope as the sample compartment,
use the Detector parameter on the Bench tab of the Experiment
Setup dialog box to select the detector you wish to use. With these
parameters set, the infrared beam is directed to the detector you
specified.
See
the OMNIC Help system for more information about setting
Experiment parameters.
Note
Before you can collect data using either the array or a single element
detector, you will need to cool it.
See
“Cooling the detector” in the “Preparing the Microscope”
chapter for instructions for cooling a detector.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...