Continuµm User Guide 151
Troubleshooting
This chapter provides information about symptoms and possible
solutions for problems you might encounter while using the
microscope. It also provides information about how to check the
microscope performance.
Symptoms
and solutions
The following table describes some problems that could occur when
you use the microscope and explains how to solve them.
If you still cannot solve the problem after using the information in
this table, contact technical support.
Symptom
Possible Cause
Solution
No signal appears in the
Bench tab live display
available through Experiment
Setup
Microscope is not in correct
mode.
Select mode by setting Sample compartment
parameter.
Find
“Sample Compartment
parameter” in OMNIC Help system Index
and go to “Specifying the sample location.”
Microscope is not focused.
Focus the microscope.
See
“Preparing the
optics” in the “Preparing the Microscope”
chapter.
Parameter settings are
incorrect.
Make sure Detector, Aperture, Velocity and
Gain parameters are set correctly.
Find
“parameters” in OMNIC Help system Index
and go to “Bench features.”
continued…
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...