Continuµm User Guide
63
Performing a
transmission
experiment
A standard transmission experiment sample was shipped in the
accessory box. This sample can be used to perform your first
transmission discrete point experiment and whenever you wish to
confirm that your microscope is operating properly. Before you
begin, be sure that the condenser is centered and focused. Data taken
through a condenser that is not centered and focused will not be
accurate or repeatable.
1. Locate the 1.0 mil PE Microscope sample slide.
It was provided with your microscope in the accessory box. The
slide includes an open beam position and an ink on polyethylene
sample.
2. Install the sample slide in the universal holder and then
install the holder on the stage.
3. If you are using a Reflachromat objective and condenser, set
the compensation rings.
Since this sample is free standing, set the rings to 0.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...