Notice
Excessive pressure can damage the ATR crystal. Be sure to stop
raising the stage in the next step when the upper indicator
illuminates.
11. Slowly raise the stage and stop immediately when the upper
Contact Alert indicator on the front panel illuminates.
This red indicator shows that the contact pressure of the sample
area against the ATR crystal is optimal for infrared analysis.
12. If you are using an ATR objective with a ZnSe or diamond
crystal, look through the viewer and make sure the sample
area of interest is positioned as desired in the field of view.
Good contact between the crystal and sample is indicated by a
darkening or “wetting” of the sample image.
If the area of interest is not positioned as desired, lower the stage
until the sample is slightly below the objective, reposition and
focus the sample with the selector in survey position, and then
return to step 10.
13. Choose Experiment Setup from the Collect menu, and verify
that the detector signal intensity is adequate.
The detector signal is shown in the live display on the Bench tab
of the Experiment Setup dialog box.
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...