Symptom
Possible Cause
Solution
No signal appears in the
Bench tab live display
available through Experiment
Setup,
continued…
Detector is not cool.
Make sure detector dewar if filled with
liquid nitrogen. Allow enough time for
detector to cool (about 10 minutes).
See
“Cooling the detector” in the “Preparing the
Microscope” chapter.
A cable is disconnected or
damaged.
Make sure the cables that connect the
microscope to the spectrometer are firmly
seated at both ends.
Replace any damaged cables.
Beam path through microscope
is blocked.
Remove any foreign objects from beam path.
Make sure Sample Compartment parameter
is set for microscope location.
Reflex aperture is closed.
Adjust aperture.
See
“Adjusting the aperture
size, shape and orientation” in the “More
about the Hardware” chapter.
If aperture does not respond, turn
microscope, computer, stage controller (if
installed), and spectrometer power off and
then power back on in the correct sequence.
See
“Turning on the microscope power” in
the “Preparing the Microscope” chapter if
you are not familiar with the power up
sequence.
Microscope is misaligned.
Center and focus the condenser and focus the
objective.
See
“Preparing the optics” in the
“Preparing the Microscope” chapter.
When you are finished, run the Noise Test in
transmission modes
See
“Testing microscope
performance” later in this chapter if you
need help.
If the image cannot be brought into sharp
focus, performance has dropped off, or the
Noise Test fails, contact technical support.
continued…
Содержание Nicolet Continuum
Страница 1: ......
Страница 9: ...Continuµm User Guide 5 Front panel ...
Страница 10: ...Thermo Scientific 6 Control panel on left side ...
Страница 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Страница 48: ...Thermo Scientific 44 ...
Страница 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Страница 80: ...Thermo Scientific 76 17 When data collection finishes use the viewer pane to isolate peaks of interest ...
Страница 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Страница 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Страница 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Страница 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Страница 126: ...Thermo Scientific 122 ...
Страница 151: ...Continuµm User Guide 147 5 Increase the reflection illumination so that it is brighter than normal viewing intensity ...
Страница 154: ...Thermo Scientific 150 ...
Страница 176: ...Thermo Scientific 172 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...
Страница 183: ...Continuµm User Guide 179 5 Choose Experiment Setup from the Collect menu The Experiment Setup dialog box appears ...